General | | |
Function | Ultrasound measurement device |
Operable with | FlexID2, LUBESense1, RS2T, RS2N, AIRSense, ULTRASense, ParaDish2, RS2A, ExtUsAmp (cable for external transducers only, since v 3.1.508) |
Measurement interface | 1 channel via a 7 pole LEMO connector |
Maximum cable length | m (ft) | Up to 30 (98) (since v 3.1.535) |
Sensitivity | Class I exceeding ASTM 1002-11 requirements for gas leak detection with the appropriate sensor. |
Keyboard | 5 function keys |
Measuring range | dBµVRMS | -6 to 99.9 (reference 0 dB = 1 µV) |
Resolution | digits | 0.1 |
Measurement bandwidth | kHz | 35 to 42 |
Signal amplification | dB | +30 to +102 by step of 6 dB |
RMS period time | 250 ms (main screen) / 3 sec (bar graph) |
Sampling frequency | kHz/ksps | 64, post-heterodyne |
ADC Resolution | bits | 12 |
Firmware update | Via SDT Updater https://sdtultrasound.com/support/software/ |
Display | 160x128 pixels Color OLED |
Environmental | | |
Operating temperature range | °C (°F) | -10 to +50 (14 to 122) non-condensing |
IP rating | IP 42 |
Compliance | EMC compliant (directive 2014/30/EU) ROHS compliant (directive 2011/65/EU) |
Standards | EN 61326-1:2013, EN 55011:2016 + A1:2017, EN 61000-4-2:2009, EN 61000-4-3:2006 + A1:2008 + IS1:2009 +A2:2010 |
Mechanical | | |
Housing material | ABS |
Dimensions Housing | mm (in) | 158x59x38.5 (6.22x2.32x1.51) |
Weight | g (oz) | 164 (5.78) |
Output interface | USB Mini-B 5-pin/App/Web 6.35 mm audio jack (see SDT LEAKReporter app) |
Power | | |
Battery | 2 AA batteries |
Autonomy* | 7 hours |
Audio | | |
Operable with | provided headset only (Peltor 3M) |
Maximum audio output (limitation) | dB SPL | +83 with the provided headset |
Headset (PPE) | 25 dB NRR |
Warranty | | |
Lifetime warranty | Visit www.sdtultrasound.com for details |